MICRO POWDER DIFFRACTION ON XTALAB SYNERGY SINGLE CRYSTAL DIFFRACTOMETERS

Introduction

Powder diffraction experiments are traditionally carried out with a dedicated powder diffractometer. These are designed for the use of relatively large amounts of sample (>>20 mg). The larger line-focused beam probes the sample's bulk nature by illuminating a large volume of material. However, if only micrograms of the sample are available, the sample volume is too small to use standard powder diffractometers. With special optics and sample holders, powder microdiffraction is viable -- but not every laboratory is equipped with them or is able to change the system configuration easily. The microfocus high-briliance X-ray sources and high-performance detectors used in Rigaku Oxford Diffraction XtaLAB Synergy systems enable users to collect microdiffraction data collection times and quality on par with dedicated powder diffractometers.

XtaLAB Synergy single crystal diffractometers can even exceed the scope of usual powder instruments. They are regularly equipped with non-ambient temperature devices. The recording of high-quality powder diffraction data is furthermore supported by the automatic divergence control, accessible only for XtaLAB Synergy systems. 

The versatile CrysAlisPro software contains all necessary plugins for powder measurements, processing, and evaluation. By utilizing CrysAlisPro and XtaLAB Synergy systems, single crystal dedicated laboratories can easily extend their range of samples to the polycrystalline phases.

Drug development is one typical use case for measurement on micrograms of powder. Therefore, the characterization of Active Pharmaceutical Ingredients (APIs) is featured in this study. Such analysis can easily be adopted for research groups involved in the development of new API or functional materials. 

Figure 1: Standard powder sample on Ø 20 mm sample holder and microscopic powder sample on Ø 0.2 mm single crystal mount

A quick phase identification of organic API with XtaLAB Synergy-S (Cu/Mo) system

The phase identification of organic polycrystalline materials can be achieved in a matter of minutes, without any time-consuming sample preparation or mounting techniques, with the help of the XtaLAB Synergy-S equipped with a high-brilliance PhotonJet Cu Kα microfocus source and HyPix 6000 detector (Fig. 2 and 3).

Figure 2: Micro powder diffraction workflow with XtaLAB Synergy-S (Cu/Mo) and CrysAlisPro

Figure 3: Extracted 1D experimental diffraction pattern of naproxen overlayed with calculated pattern based on PDF4-Organics entry 20642110 in CrysAlisPro

Figure 4: Powder samples on Kapton® and nylon loops and in quartz and Kapton® capillaries

The ease of sample preparation

The preparation of powder samples on single-crystal mounts is very easy and straightforward. Using a drop of oil, the powder sample sticks to the loops. Alternatively, quartz or Kapton® capillaries can be used as shown in Fig. 4. The samples can be mounted on magnetic bases or standard mounts and easily centered in the beam, just like a single crystal. 

Rigaku’s HyPix technology advantage

XtaLAB Synergy systems come equipped with Rigaku HyPix detectors as standard. Hybrid X-ray Photon Counting (HPC) detectors are considered high-end for dedicated powder diffractometers, e.g., the SmartLab (Fig. 5). The benefits, such as high dynamic range, single-photon sensitivity, excellent spatial resolution, zero noise, and single-pixel point spread function are well proven and tested by the powder and thin layer XRD community. 

HyPix-3000_picture_Ver1_2021.02.05HyPix-Arc 150°_PH_2020.11.2

Figure 5: A SmartLab’s HyPix-3000 next to XtaLAB Synergy’s HyPix-Arc 150º, both using HPC technology

Continuous variable divergence control improves peak resolution

Microfocus sources, by design, use convergent beams to increase flux density. This means the beam diverges after the sample, which can be undesirable. Though HyPix detectors have an excellent spatial resolution suitable for quick phase identification at short detector distances, for complex mixtures or low symmetry materials this can result in peak overlap. Longer detector distances may improve resolution but with limited success. PhotonJet Cu sources can be equipped with a divergence control slit as an option. A simple software slider in CrysAlisPro allows the divergence to be adjusted down to 1 mRad, which can dramatically improve peak shape and thus data quality as shown in Fig. 6

Hardware solution for sample quality issues

Good sample preparation is crucial in powder measurement and analysis. Proper sample grinding and randomization of the crystallite orientation is even more important for small samples. But in the case of some organic materials grinding can lead to amorphization or is simply not possible due to limited sample availability.
To minimize the influence of preferred orientation, the Gandolfi Method can be easily applied with XtaLAB Synergy Universal Goniometer.

Figure 6: Beam divergence improves peak resolution. A broad, ill-defined peak with the raw beam (blue) is resolved into 2 peaks when using low divergence (red)

Data collection with Gandolfi scans randomizes the sample orientation in the beam by driving both phi and omega circles at the same time. If required, the scan can be repeated for different kappa settings. The calculation of possible scans and data integration is done automatically by CrysAlisPro software. Some exemplary results from this procedure are shown in Fig. 7

Figure 7: Application of Gandolfi scans on flufenamic acid sample. (XtaLAB Synergy-S(Cu/Mo) detector distance 60 mm, exposure time 60 s per scan)

Software advantages of CrysAlisPro

Constant improvement, inspired by users, led to the development and inclusion of dedicated powder data processing tools in CrysAlisPro. These tools allow for the extraction of high-quality 1D powder diffraction patterns from 2D diffraction images. The algorithm accounts for instrument model adjustments and provides background removal, baseline correction options, and multiple modes of pattern extraction. Quick phase match with known structure factors as a graphical representation of measured and literature patterns are fully implemented within the software.

New phase or just a mixture?

A quick screening of a sample revealed additional intensities next to the expected pattern (Fig. 8). Overlaying peak prediction markers for a suspected contaminant within CrysAlisPro allows for quick confirmation of their presence. An extended measurement of the same mixture in Fig. 9 allows for Rietveld refinement in dedicated programs.

Figure 8: Screening flufenamic acid - naproxen mixture on an XtaLAB Synergy-S Cu at 190 mm, 2.9 mRad, 180 s

Figure 9: Rietveld refinement indicates a 2:1 flufenamic acid: naproxen mixture on an XtaLAB Synergy-S Cu at 190 mm, 1.0 mRad, 1 h)

Multiple sample screening and batch-processing

Quick and easy sample preparation, as well as a large detector, allow for fast screening of multiple samples. This is further supported in CrysAlisPro with a batch-processing feature. The fingerprint-region of six APIs, shown in Fig. 10, were measured within one minute each and processed swiftly as one batch with the same prerequisites.

Figure 10: Screening of six APIs XtaLAB Synergy-S Cu at 120 mm, 3.3 mRad, 60 s each)

Additional benefits of using XtaLAB Synergy systems

• Quick and easy sample centering with a video camera on the system
• Variable temperature studies with existing cryo-device
• XRF suppression with any HyPix detector.
• Switch radiation source with DW or dual-source systems.
• Employ XtalCheck-S for automated powder screening
• PDF analysis accessible due to versatile goniometer and photon-counting detector.

Conclusions

XtaLAB Synergy single crystal diffractometers are excellent instruments for collecting powder microdiffraction data from very small amounts of organic API materials. A XtaLAB Synergy diffractometer and CrysAlisPro allow for quick identification of mixtures and new crystalline phases.